FDA Adverse Event
Malfunction
Summary report: N
9990
MDR report key: 3161696
·
Received June 3, 2013
Report
- Report Number
- 1720753-2013-06722
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 23, 2013
- Report Date
- June 4, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HAND SWITCH WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN X-RAY SWITCH STUCK ERROR MESSAGE. THE ERROR MESSAGE COULD NOT BE CLEANED AND THE FLUOROSCOPY FUNCTION WAS DISABLED AS A RESULT. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243629 | 9990 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |