FDA Adverse Event
Malfunction
Summary report: N
TRACTION MACHINE - ME 4000
MDR report key: 3161615
·
Received June 5, 2013
Report
- Report Number
- MW5030472
- Event Type
- Malfunction
- Date Received
- June 5, 2013
- Date of Event
- May 17, 2013
- Report Date
- May 20, 2013
- Manufacturer
- METTLER ELECTRONICS CORP
- Product Code
- ITH
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
A PT WAS PLACED INTO A HARNESS FOR CERVICAL TRACTION. THE CABLE WAS PRETESTED FOR SLIPPING BY PT. THE CABLE WAS ATTACHED TO THE HARNESS BY THE PT. AS THE CABLE WAS ATTACHED, THERE WAS NO NOTED ABNORMALITY OR FRAYING OF THE CABLE. THE TRACTION PULL WAS INITIATED. WHEN THE PULL REACHED ABOUT 17 LBS ON THE DIGITAL READOUT, THE CABLE BROKE FROM THE CLIP. THE TRACTION HARNESS WAS DISENGAGED FROM THE MOTOR SUDDENLY. 140291-2013-8.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 247842 | TRACTION MACHINE - ME 4000 | MTD - ME 4000 | ITH | METTLER ELECTRONICS CORP | ME 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |