NOVASURE IMPEDANCE CONTROLLED EA SYSTEM
Report
- Report Number
- 1222780-2013-00105
- Event Type
- Injury
- Date Received
- May 7, 2013
- Date of Event
- April 24, 2013
- Report Date
- May 10, 2013
- Manufacturer
- HOLOGIC
- Product Code
- MNB
- PMA / PMN Number
- P010013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE HAS NOT YET BEEN RETURNED THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. IF THE DEVICE IS RETURNED AND EVALUATION COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).
IT WAS REPORTED THAT UPON SEATING THE DISPOSABLE NOVASURE DEVICE DURING AN ENDOMETRIAL ABLATION THE PHYSICIAN "DIDN'T FEEL LIKE HE WAS SEATING [THE DEVICE] CORRECTLY." A HYSTEROSCOPY WAS DONE AND HE "SAW A PERFORATION AT THE FUNDUS." THE ATTEMPTED NOVASURE PROCEDURE WAS ABORTED. NO TREATMENT WAS NEEDED AND THE PATIENT WAS DISCHARGED HOME. A HYSTEROSCOPY, DILATATION AND CURETTAGE (D&C), AND SOUNDING (NOT HOLOGIC DEVICES) WERE PERFORMED PRIOR TO THE ATTEMPTED ABLATION. IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 198176 | NOVASURE IMPEDANCE CONTROLLED EA SYSTEM | MNB | HOLOGIC | NS2000 | 12K10RA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other | RADIO FREQUENCY CONTROLLER: SERIAL #UNK |