FDA Adverse Event Injury Summary report: N

NOVASURE IMPEDANCE CONTROLLED EA SYSTEM

MDR report key: 3161579 · Received May 7, 2013

Report

Report Number
1222780-2013-00105
Event Type
Injury
Date Received
May 7, 2013
Date of Event
April 24, 2013
Report Date
May 10, 2013
Manufacturer
HOLOGIC
Product Code
MNB
PMA / PMN Number
P010013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. IF THE DEVICE IS RETURNED AND EVALUATION COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT UPON SEATING THE DISPOSABLE NOVASURE DEVICE DURING AN ENDOMETRIAL ABLATION THE PHYSICIAN "DIDN'T FEEL LIKE HE WAS SEATING [THE DEVICE] CORRECTLY." A HYSTEROSCOPY WAS DONE AND HE "SAW A PERFORATION AT THE FUNDUS." THE ATTEMPTED NOVASURE PROCEDURE WAS ABORTED. NO TREATMENT WAS NEEDED AND THE PATIENT WAS DISCHARGED HOME. A HYSTEROSCOPY, DILATATION AND CURETTAGE (D&C), AND SOUNDING (NOT HOLOGIC DEVICES) WERE PERFORMED PRIOR TO THE ATTEMPTED ABLATION. IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
198176 NOVASURE IMPEDANCE CONTROLLED EA SYSTEM MNB HOLOGIC NS2000 12K10RA

Patients

Seq Age Sex Outcome Treatment
1 UNK Other RADIO FREQUENCY CONTROLLER: SERIAL #UNK