FDA Adverse Event Malfunction Summary report: N

STERLING?

MDR report key: 3161560 · Received June 11, 2013

Report

Report Number
2134265-2013-03888
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 16, 2013
Report Date
May 16, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053118
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MANUFACTURER: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PERCUTANEOUS PERIPHERAL INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS SUPERFICIAL FEMORAL ARTERY. A 6.0MM X 20MM X 135CM (4F) STERLING BALLOON CATHETER WAS USED TO TREAT THE LESION. UPON THE FIRST INFLATION AT 7 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
263280 STERLING? CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939031602010 14356479

Patients

Seq Age Sex Outcome Treatment
1 GUIDE WIRE: CRUISE/ SJM| INTORDUCER SHEATH: MEDIKIT 6F