FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3161555 · Received June 11, 2013

Report

Report Number
1416980-2013-14913
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 17, 2013
Report Date
May 17, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE. THE DEVICE IS REPORTED TO BE AVAILABLE FOR EVALUATION. SHOULD THE DEVICE BE RETURNED OR ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RECEIVED AND AN EVALUATION WAS PERFORMED ON THE DEVICE. VISUAL INSPECTION, LEAK TESTING, AND CLEAR PASSAGE TESTING WAS PERFORMED WITH NO ISSUES NOTED. ANALYSIS OF THE DEVICE WAS UNABLE TO CONFIRM THE REPORTED PARTICULATE MATTER. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE TROUBLESHOOTING FOR AN UNRELATED ALARM ON THE HOME CHOICE (HC) MACHINE, THE HOME PATIENT (HP) NOTICED SOMETHING BLACK IN THE PATIENT LINE DURING FILL 1 OF 8, PATIENT CONNECTED. DURING THE FOLLOW UP WITH THE PERITONEAL DIALYSIS REGISTERED NURSE (PDRN), IT WAS REPORTED THAT THERE WAS NOTHING BLACK IN THE PATIENT LINE AND THAT THE TUBING WAS POSITIONED ON SOMETHING THAT MADE IT LOOK BLACK. HOWEVER, FURTHER CLARIFICATION WITH THE HP REVEALED THAT THE HP DID SEE SOMETHING VERY SMALL AND BLACK IN THE LINE. THE HP STATED THAT WHEN THEY SQUEEZED THE LINE, THEY SAW IT MOVE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262324 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME H13B23086

Patients

Seq Age Sex Outcome Treatment
1 65 YR HOMECHOICE