CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-04709
- Event Type
- Injury
- Date Received
- June 5, 2013
- Date of Event
- May 14, 2013
- Report Date
- May 14, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVAL RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-04708. IT WAS REPORTED THE PT WAS EXPERIENCING HEATING AT THE IPG SITE WHILE RECHARGING. THE PT WAS CONTACTED AND IT WAS REPORTED THERE HAD BEEN NO SKIN BURNS NOR MARKS ON THE SKIN. THE PT WAS EXPERIENCING UNCOMFORTABLE HEATING AFTER ABOUT 15 MINS USING THE CHARGING SYSTEM. THE PT REPORTED SHE DID USE THE ANTENNA POUCH AND ALSO PLACED A CLOTH BETWEEN THE ANTENNA AND THE SKIN. IT WAS REPORTED THE PT WAS RECHARGING WITH THE SYSTEM TURNED OFF. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 248286 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | 2889950 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Required Intervention | IMPLANT DATE:| SCS ANCHOR: MODEL 1194 (2)| SCS LEAD: MODEL 3214| IMPLANT DATE: |