EON MINI
Report
- Report Number
- 1627487-2013-01641
- Event Type
- Injury
- Date Received
- June 5, 2013
- Date of Event
- May 13, 2013
- Report Date
- May 13, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS ASSOCIATED WITH A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-1642. IT WAS REPORTED THE PATIENT WAS EXPERIENCING AN UNCOMFORTABLE HEATING SENSATION AT HIS IPG SITE WHILE CHARGING. A NEW LE CHARGER WAS SENT TO THE PATIENT TO ADDRESS THE ISSUE. FOLLOW UP INFORMATION IDENTIFIED THE NEW LE CHARGER RESOLVED THE ISSUE. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 249196 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3712111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention | IMPLANT DATE:| SCS LEAD: MODEL 3186 (2)| SCS ANCHOR: MODEL 1192 (2)| IMPLANT DATE: |