FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 3161500 · Received June 11, 2013

Report

Report Number
2250051-2013-00162
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 14, 2013
Report Date
June 11, 2013
Manufacturer
HAMILTON BONADUZ AG
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2013 AN OCD FIELD ENGINEER (FE) ARRIVED AT THE CUSTOMER SITE AND REPLACED TIP AND PLUNGER CLAMP IN POSITION 2. THE FE INSPECTED ALL TIP AND PLUNGER CLAMPS. THE FE TESTED THE LLD WITH THE SPLLD TEST. THE FE TESTED THE SUMMIT WITH (B)(4) TEST. ALL TEST PASSED. REPAIRS HAVE RETURNED THIS INSTRUMENT TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE THE CORRECT AMOUNT OF SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262113 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTOR MMH HAMILTON BONADUZ AG 936400

Patients

Seq Age Sex Outcome Treatment
1