ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3161480 · Received June 11, 2013

Report

Report Number
2531779-2013-08044
Event Type
Injury
Date Received
June 11, 2013
Report Date
May 13, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE METER HAS NOT BEEN REQUESTED FOR RETURN TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS METER AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. TROUBLESHOOTING OF THE EVENT INDICATED THAT THE PUMP AND METER REMOTE WERE NOT PAIRED RESULTING IN LACK OF COMMUNICATION BETWEEN DEVICES. THE PUMP CAN STILL BE MANIPULATED WITHOUT PAIRING THE PUMP WITH A METER REMOTE.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS ON (B)(6) 2013 REPORTING JUST RECEIVING A NEW PUMP BUT DID NOT PAIR THE PUMP TO THE METER REMOTE. THE REPORTER INDICATED THAT THE PUMP AND METER WERE NOT COMMUNICATING TO DELIVER BOLUSES RESULTING IN A BLOOD GLUCOSE OF 300 MG/DL WITH STOMACH PAIN. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPERGLYCEMIA RELATED TO USE OF AN UNPAIRED METER REMOTE AND PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
263214 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 49 YR Life Threatening