FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3161445 · Received June 5, 2013

Report

Report Number
1627487-2013-15771
Event Type
Injury
Date Received
June 5, 2013
Date of Event
May 15, 2013
Report Date
May 15, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-15770. THE PATIENT REPORTED SHE EXPERIENCES A HEATING SENSATION ALL THE TIME. THE PATIENT STATED THE HEATING SENSATION IS WORSE WHEN SHE CHARGES HER IPG. ADDITIONALLY, THE PATIENT STATES SHE HAS LOST WEIGHT AND HER IPG STICKS OUT AND CATCHES ON HER CLOTHES. A REPLACEMENT LE CHARGING SYSTEM WAS SENT TO THE PATIENT. THE PATIENT IS TO MEET WITH HER PHYSICIAN AND THE SJM REPRESENTATIVE AS THE NEXT COURSE OF ACTION. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249039 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3350616

Patients

Seq Age Sex Outcome Treatment
1 38 YR Required Intervention IMPLANT DATE:| SCS ANCHOR, MODEL 1192 (2)| IMPLANT DATE:| SCS LEAD, MODEL 3186 (2)