FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3161190 · Received June 11, 2013

Report

Report Number
3008382007-2013-16569
Event Type
Malfunction
Date Received
June 11, 2013
Report Date
June 7, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Removal / Correction Number
2013-IE-017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATELY HIGH RESULTS. READINGS OF 176MG/DL ON THE SUBJECT METER AND 120MG/DL ON AN UNKNOWN METER WERE TAKEN WITHIN 30 MINUTES OF EACH OTHER. THIS ISSUE IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S ACCURACY CRITERIA OF <30% DIFFERENCE COMPARED TO RESULTS TAKEN ON ANOTHER METER WITHIN 30 MINUTES OF EACH OTHER. THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262603 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3418735

Patients

Seq Age Sex Outcome Treatment
1