FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3160921 · Received June 11, 2013

Report

Report Number
3008382007-2013-16436
Event Type
Malfunction
Date Received
June 11, 2013
Report Date
June 6, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Removal / Correction Number
2013-IE-017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE RESULTS. RESULTS OF 72MG/DL AND 97MG/DL WERE OBTAINED ON THE SAME LIFESCAN METER WITHIN 20 MINUTES OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S PRECISION CRITERIA OF <20% DIFFERENCE BETWEEN RESULTS AVERAGING >75MG/DL PERFORMED ON THE SAME METER WITHIN 20 MINUTES. THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262179 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1