FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 3160875 · Received June 11, 2013

Report

Report Number
3008382007-2013-16438
Event Type
Malfunction
Date Received
June 11, 2013
Report Date
June 6, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K082590
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

(B)(4) - PRODUCT EVALUATION: THE SUBJECT METER WAS RETURNED ON (B)(4) 2013 AND ANALYSIS COMPLETED ON (B)(4) 2013 BY LIFESCAN PRODUCT ANALYSIS. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. THE DEVICE MET ALL SPECIFICATIONS FOR TESTING AND NO ISSUES WERE OBSERVED DURING INVESTIGATION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING DOES NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262948 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3292998

Patients

Seq Age Sex Outcome Treatment
1 64 YR