HOMECHOICE PRO
Report
- Report Number
- 1416980-2013-14910
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Date of Event
- May 17, 2013
- Report Date
- May 17, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE IS REPORTED TO BE AVAILABLE BUT HAS NOT BEEN RECEIVED AT THIS TIME. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). A DEVICE HISTORY REVIEW WAS PERFORMED FINDING NO EXCEPTION, NONCONFORMANCE, OR REWORK THAT OCCURRED DURING THE MANUFACTURING OF THE DEVICE WITH THIS SERIAL NUMBER. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS THAT WERE RELATED TO THE REPORTED CONDITION. DURING EVALUATION OF THE DEVICE SIMULATED THERAPY WAS PERFORMED AND COMPLETED WITH NO ISSUES NOTED. THE REPORTED ISSUE COULD NOT BE CONFIRMED IN THE LOG REVIEW NOR DUPLICATED DURING TESTING OF THE DEVICE. DURING VISUAL INSPECTION OF THE DEVICE, NO ISSUES WERE FOUND. THE ROOT CAUSE OF THE REPORTED ISSUE CANNOT BE DETERMINED BASED ON THE INFORMATION AVAILABLE. THE DEVICE WAS SENT FOR NORMAL SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE HOMECHOICE (HC) MACHINE DID NOT DISPLAY THE CORRECT FILL AND DRAIN VOLUMES. THE HOME PATIENT'S (HP) PRESCRIPTION WAS FOR APPROXIMATELY 16 L WHILE THE VOLUMES RECORDED ON THE MACHINE AT THE END OF THERAPY WERE 14 L FILLED AND 15 L DRAINED. AT THE END OF THE THERAPY, ALL OF THE BAGS WERE EMPTY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263207 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN PD SOLUTION| 5L AND 2L BAGS |