UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2013-00403
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- May 15, 2013
- Report Date
- May 15, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
QUALITY CONTROL (QC) ON THE DAY OF THE EVENT HAD BEEN EXHIBITING SOME IMPRECISION, WITH HIGH OUT OF RANGE RECOVERY PRIOR TO THE EVENT. THE SYSTEM WAS RECALIBRATED AND QC RECOVERED NEAR THE MEAN, AT WHICH THE TIME THE SAMPLES WERE RUN. NO SAMPLE COLLECTION OR METHOD OF ANALYSIS WAS PROVIDED BY THE CUSTOMER. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE DISASSEMBLED AND CLEANED THE FLOWCELL. THE FSE OBSERVED BUILD UP BY BOTH THE NA AND CL PORTS. THE FSE REPLACED THE NA REFERENCE ELECTRODE, CL ELECTRODE TIP AND BOTH CO2 ELECTRODES. THE FSE ALSO REPLACED THE CARBON BRIDGE. THE FSE ALSO REPLACED THE RATIO PUMP AS A PREVENTIVE MAINTENANCE SINCE IT WAS DUE TO BE REPLACED IN TWO WEEKS. FAILURE MODE IS UNKNOWN. MULTIPLE PARTS WERE REPLACED AND ACTIONS TAKEN, ANY OF WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE EVENT. MDR 2050012-2013-00402 IS ASSOCIATED WITH THIS EVENT AND DOCUMENTS THE RESULTS OBTAINED ON (B)(6) 2013.
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER GENERATED FALSE HIGH SODIUM (NA), POTASSIUM (K), CHLORIDE (CL), AND/OR CARBON DIOXIDE (CO2) RESULTS FOR THREE (3) PATIENT SAMPLES ON TWO (2) DIFFERENT DAYS ((B)(6) 2013). THIS REPORT DOCUMENTS THE PATIENT RESULTS GENERATED ON (B)(6) 2013 FOR ONE (1) OF THE THREE PATIENT SAMPLES THAT WERE AFFECTED. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. WHEN THE ISSUE WAS NOTICED, THE SAMPLES WERE REPEATED ON AN ALTERNATE INSTRUMENT AND YIELDED RESULTS WITHIN THE NORMAL RANGE AND THOSE RESULTS WERE AMENDED. THE CUSTOMER HAS CONFIRMED THAT THERE HAS BEEN NO EFFECT TO PATIENTS OR CHANGE IN PATIENT TREATMENT IN CONNECTION TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260093 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |