FDA Adverse Event Malfunction Summary report: N

THORACIC GRASPER INSTRUMENT

MDR report key: 3160747 · Received June 10, 2013

Report

Report Number
2955842-2013-02102
Event Type
Malfunction
Date Received
June 10, 2013
Date of Event
May 16, 2013
Report Date
May 17, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE BLACK TUBE INSULATION WAS DAMAGED AT THE DISTAL END. THE INSULATION WAS GOUGED ON ONE SIDE AND HAD A .100 X .080 PIECE MISSING ROUGHLY 4 ABOVE THE SNAKE WRIST. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING THE DA VINCI SI SURGICAL THORACIC GRASPER INSTRUMENT WAS NOTED TO HAVE A CUT IN THE INSULATION ON THE SHAFT OF THE INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260776 THORACIC GRASPER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420343-01 S10130214 108

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES