FDA Adverse Event
Malfunction
Summary report: N
THORACIC GRASPER INSTRUMENT
MDR report key: 3160747
·
Received June 10, 2013
Report
- Report Number
- 2955842-2013-02102
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- May 16, 2013
- Report Date
- May 17, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE BLACK TUBE INSULATION WAS DAMAGED AT THE DISTAL END. THE INSULATION WAS GOUGED ON ONE SIDE AND HAD A .100 X .080 PIECE MISSING ROUGHLY 4 ABOVE THE SNAKE WRIST. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING REPROCESSING THE DA VINCI SI SURGICAL THORACIC GRASPER INSTRUMENT WAS NOTED TO HAVE A CUT IN THE INSULATION ON THE SHAFT OF THE INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260776 | THORACIC GRASPER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420343-01 | S10130214 108 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES |