FDA Adverse Event
Malfunction
Summary report: N
RESERVOIR 3ML
MDR report key: 3160735
·
Received June 10, 2013
Report
- Report Number
- 3004209178-2013-94617
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- May 28, 2013
- Report Date
- May 28, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
INSPECTED THREE OPENED AND USED RESERVOIRS. PRE-FILL RESERVOIR TEST PERFORMED. ONE RESERVOIRS FAILED PER SPECIFICATION. RESERVOIR AND TRANSFER GUARDS OCCLUDED. THE TWO REMAINING RESERVOIRS PASSED PER INSPECTION. NO OCCLUDED OR AIR BUBBLES ANOMALIES WERE OBSERVED ANALYSIS.
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED PAST THE O-RINGS DURING THE FILLING PROCESS. THE BLOOD GLUCOSE READING WAS 76MG/DL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 261149 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A | H8652178 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |