FDA Adverse Event
Malfunction
Summary report: N
MICRO SAGITTAL SAW
MDR report key: 3160703
·
Received June 10, 2013
Report
- Report Number
- 0001811755-2013-01330
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- May 29, 2013
- Report Date
- May 29, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND TO BE BUILT UP THROUGHOUT THE INSIDE OF THE HANDPIECE.
Description of Event or Problem · 1
THE MICRO SAGITTAL SAW WAS RETURNED FOR SERVICE. UPON EVALUATION AT THE MANUFACTURER IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260948 | MICRO SAGITTAL SAW | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |