ENTRUST ESCUDO VR
Report
- Report Number
- 9614453-2013-01430
- Event Type
- Injury
- Date Received
- June 10, 2013
- Report Date
- July 26, 2013
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE ACTUAL LONGEVITY WAS LESS THAN 80% OF THE 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THE DEVICE BATTERY REACHED ELECTIVE REPLACEMENT INDICATOR STATUS EARLIER THAN EXPECTED. THE DEVICE WAS REMOVED AND A NEW DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THE DEVICE BATTERY REACHED ELECTIVE REPLACEMENT INDICATOR STATUS EARLIER THAN EXPECTED. THE DEVICE WAS REMOVED AND A NEW DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260183 | ENTRUST ESCUDO VR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | IPG MFG SWITZERLAND | D144VRC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |