FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3160487 · Received June 10, 2013

Report

Report Number
2955842-2013-02091
Event Type
Malfunction
Date Received
June 10, 2013
Date of Event
May 15, 2013
Report Date
June 28, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K072627
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATIONS CONFIRMED THERE WAS BROKEN CONDUCTOR WIRE AT THE YAW PULLEY EXIT. THE WIRE WAS DETACHED FROM ITS CONNECTION AT THE GRIP. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND FAILED. THE YAW PULLEY SHOWED NO SIGNS OF ARCING. NO OTHER DAMAGE WAS FOUND. RECURRENCE OF THE ALLEGED FAILURE MODE IS NOT EXPECTED TO CAUSE OR CONTRIBUTE TO A DEATH, SERIOUS INJURY OR SERIOUS DETERIORATION IN THE STATE OF HEALTH OF A PATIENT. BASED ON THIS ADDITIONAL FAILURE ANALYSIS INVESTIGATION INFORMATION, THIS MDR REPORT IS BEING RETRACTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, A CABLE ON THE PK DISSECTING FORCEPS INSTRUMENT BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259724 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 M10130118 533

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS, ACCESSORIES, ESU