PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-02091
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- May 15, 2013
- Report Date
- June 28, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K072627
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATIONS CONFIRMED THERE WAS BROKEN CONDUCTOR WIRE AT THE YAW PULLEY EXIT. THE WIRE WAS DETACHED FROM ITS CONNECTION AT THE GRIP. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND FAILED. THE YAW PULLEY SHOWED NO SIGNS OF ARCING. NO OTHER DAMAGE WAS FOUND. RECURRENCE OF THE ALLEGED FAILURE MODE IS NOT EXPECTED TO CAUSE OR CONTRIBUTE TO A DEATH, SERIOUS INJURY OR SERIOUS DETERIORATION IN THE STATE OF HEALTH OF A PATIENT. BASED ON THIS ADDITIONAL FAILURE ANALYSIS INVESTIGATION INFORMATION, THIS MDR REPORT IS BEING RETRACTED.
IT WAS REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, A CABLE ON THE PK DISSECTING FORCEPS INSTRUMENT BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 259724 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10130118 533 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS, ACCESSORIES, ESU |