CARELINK
Report
- Report Number
- 2182208-2013-01830
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Report Date
- April 30, 2013
- Manufacturer
- RICE CREEK MFG
- Product Code
- DXY
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE BACKING WAS MISSING ON THE LABEL. IT WAS ALSO NOTED THAT THE DEVICE FAILED TELEMETRY TESTING DUE TO THE CABLE BEING OUT OF ELECTRICAL SPECIFICATION. CONCOMITANT PRODUCT: PRODUCT ID 2090 PROGRAMMER. (B)(4).
IT WAS REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS MISSING ITS TRACTION PAD. THE PROGRAMMER HEAD WAS RETURNED FOR SERVICE. THE PROGRAMMER HEAD WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 259762 | CARELINK | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | RICE CREEK MFG | 2067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |