FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3160348 · Received June 10, 2013

Report

Report Number
2182208-2013-01830
Event Type
Malfunction
Date Received
June 10, 2013
Report Date
April 30, 2013
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE BACKING WAS MISSING ON THE LABEL. IT WAS ALSO NOTED THAT THE DEVICE FAILED TELEMETRY TESTING DUE TO THE CABLE BEING OUT OF ELECTRICAL SPECIFICATION. CONCOMITANT PRODUCT: PRODUCT ID 2090 PROGRAMMER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS MISSING ITS TRACTION PAD. THE PROGRAMMER HEAD WAS RETURNED FOR SERVICE. THE PROGRAMMER HEAD WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259762 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2067

Patients

Seq Age Sex Outcome Treatment
1