FDA Adverse Event Injury Summary report: N

CAPSUREFIXNOVUS

MDR report key: 3160318 · Received June 10, 2013

Report

Report Number
2649622-2013-07945
Event Type
Injury
Date Received
June 10, 2013
Date of Event
March 22, 2013
Report Date
March 22, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD, THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 5076 IMPLANTABLE PACING LEAD (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT SHORTLY AFTER IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD HAD NO VENTRICULAR CAPTURE. A LEAD DISLODGEMENT WAS CONFIRMED AND REPOSITIONING WAS ATTEMPTED, BUT THE LEAD COULD NOT ADVANCE DUE TO PATIENT ANATOMY. THE LEAD WAS EXPLANTED AND REPLACED WITH A LONGER LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259737 CAPSUREFIXNOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00080 YR Hospitalization| R VEDR01 IMPLANTABLE PULSE GENERATOR (IPG)