ATTAIN STARFIX
Report
- Report Number
- 2649622-2013-07946
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- August 11, 2012
- Report Date
- March 15, 2013
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P060039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED; THE TIP SEAL ON THE DISTAL ELECTRODE OF THE LEAD SHOWED EVIDENCE OF BLOOD INGRESSION, AND THE OVERLAY TUBING OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE LEAD WAS ATTEMPTED AT IMPLANT BUT WAS UNSUITABLE DUE TO PATIENT ANATOMY. THE LEAD WAS ALSO PAST ITS USE-BY -DATE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.