CONSULTA CRT-D
Report
- Report Number
- 3004209178-2013-10074
- Event Type
- Injury
- Date Received
- June 10, 2013
- Report Date
- April 24, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Removal / Correction Number
- Z-0111-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE DID NOT DETECT ANY PERFORMANCE ISSUES, BUT THE CALCULATED LONGEVITY RESULT OF 84% OF EXPECTED WAS LESS THAN THE PREDICTED VALUE BASED ON THE AVAILABLE PROGRAMMED PARAMETERS. IN THE ABSENCE OF THE COMPLETE PROGRAMMING HISTORY, IT IS NOT POSSIBLE TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: MODEL 4076-52, IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 2009; MODEL 4194-88, IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 2009; MODEL 6947-65, IMPLANTABLE TACHY LEAD, IMPLANTED: (B)(6) 2009. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE DEVICE DID NOT MEET LONGEVITY EXPECTATIONS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.