FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 3160269 · Received June 10, 2013

Report

Report Number
2649622-2013-07965
Event Type
Injury
Date Received
June 10, 2013
Report Date
March 25, 2013
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)  IMPLANTED: (B)(6) 2009, 6947 IMPLANTABLE TACHY LEAD IMPLANTED: (B)(6) 2009, 5076 IMPLANTABLE PACING LEAD IMPLANTED: (B)(6) 2009. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS EXPERIENCING HIGH THRESHOLDS AND NO CAPTURE. THE LEAD WAS TO BE EXPLANTED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, AFTER THE DEVICE WAS REMOVED, IT WAS NOTED THAT BOTH THE LV AND RIGHT VENTRICULAR (RV) LEADS HAD AN APPARENT FRACTURE, AND THAT THE LEADS WERE ALL "JUMBLED UP IN THE POCKET." THE LV AND RV LEADS WERE THEN CAPPED, AND WERE LATER EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260667 ATTAIN OTW PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 419478

Patients

Seq Age Sex Outcome Treatment
1 00061 YR Hospitalization| R