PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-02083
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- May 15, 2013
- Report Date
- May 16, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE ONE CONDUCTOR WIRE WAS BROKEN AT THE YAW PULLEY EXIT. THE WIRE WAS DETACHED FROM THE CONNECTION AT THE GRIP. THE INSTRUMENT FAILED ELECTRICAL CONTINUITY TESTING. NO SIGN OF ARCING WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE THE PK DISSECTING FORCEPS INSTRUMENT WIRE BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260484 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10130121 713 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |