ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-08031
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Report Date
- May 16, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION; HOWEVER, PRODUCT INVESTIGATION HAS NOT BEEN COMPLETED. ONCE EVALUATION HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 08/06/2013 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2013 WITH THE FOLLOWING FINDINGS:DURING TESTING, THE DISPLAY SCREEN WAS FOUND TO BE DIM/FADED. A TEST SCREEN WAS INSERTED AND WAS FOUND TO FUNCTION PROPERLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING) ISSUE. THE DISPLAY ON THE ANIMAS PUMP REPORTEDLY IS HARD TO READ. THERE IS NO INDICATION THAT THE PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE REMAINED UNRESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260469 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |