DEXTRUS 4135
Report
- Report Number
- 1028232-2013-01642
- Event Type
- Injury
- Date Received
- June 10, 2013
- Date of Event
- April 23, 2013
- Report Date
- May 31, 2013
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVN
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. DURING THE MECHANICAL INSPECTION A STYLET COULD NOT BE PROPERLY INTRODUCED AND ADVANCED WITHIN THE LEAD'S LUMEN. THE LEAD WAS VISUAL ANALYSED AND COAGULATED BLOOD WAS IDENTIFIED IN THE LUMEN OF THE LEAD CAUSING THE INABILITY TO ADVANCE THE STYLET PROPERLY. THESE BLOOD RESIDUALS WERE MOST LIKELY INTRODUCED INTO THE LUMEN OF THE LEAD BY MEANS OF STYLETS USED DURING THE SURGERY. FURTHER VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS OF THE OUTER COIL WHICH OCCURRED MOST LIKELY DURING SURGERY. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD WAS DISLODGED. THIS RA LEAD WAS EXPLANTED, REPLACED SUCCESSFULLY AND WILL BE RETURNED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260771 | DEXTRUS 4135 | PACER LEAD | NVN | BIOTRONIK SE & CO. KG | 358744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |