FDA Adverse Event Malfunction Summary report: N

LUMENIS ONE

MDR report key: 3160112 · Received June 10, 2013

Report

Report Number
3004135191-2013-00039
Event Type
Malfunction
Date Received
June 10, 2013
Date of Event
May 23, 2013
Report Date
June 10, 2013
Manufacturer
LUMENIS, LTD.
Product Code
GEX
PMA / PMN Number
K060448
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

LUMENIS INVESTIGATED THE REPORTED COMPLAINT. A LUMENIS TECHNICAL EXPERT COMPLETED AN EXAMINATION OF THE SUBJECT DEVICE IDENTIFYING A CRACK IN A COOLANT TUBE FITTING TO BE THE CAUSE OF THE EVENT REPORT. LUMENIS IS REPORTING THIS COMPLAINT SINCE IT IS SIMILAR TO AN EVENT IN WHICH THE COOLING SYSTEM LEAK RESULTED IN COMBUSTION OF THE HIGH VOLTAGE POWER SUPPLY.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT A COOLING SYSTEM LEAK OCCURRED IN A LUMENIS ONE LASER. NO REPORT OF RELATED INJURY WAS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260981 LUMENIS ONE INTENSE PULSE LIGHT DELIVERY DEVICE GEX LUMENIS, LTD. LUMENIS ONE

Patients

Seq Age Sex Outcome Treatment
1 Other