ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-08010
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Report Date
- May 18, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 08/05/2013 WITH THE FOLLOWING RESULTS:TESTING CONFIRMED THE CONTRAST AND UP KEYS ARE INTERMITTENTLY RESPONDING TO USER PRESSES AND NEED EXCESSIVE FORCE TO ACTIVATE. REMOVED THE KEYPAD COVER AND FOUND CONTAMINATION UNDER THE UP, DOWN AND OK KEY CONTACTS. UNRELATED TO THE COMPLAINT, A PINKISH CONTRAST WAS OBSERVED ON THE DISPLAY SCREEN. REMOVED THE PUMP COVER AND REPLACED PINK DISPLAY WITH TEST DISPLAY. THE TEST DISPLAY HAS NORMAL CONTRAST WITH NO VISIBLE SIGNS OF DISCOLORATION. ALSO UNRELATED,THE BLACK BOX SHOWS THE TIME AND DATE RESETTING TO THE DEFAULT SETTING ON (B)(6) 2013 (11:49->16:02=4HRS 13 MINS.) A VISIBLE INSPECTION SHOWED THE INTERNAL BATTERY WAS LEAKING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THE UP/DOWN/OK BUTTONS WERE UNRESPONSIVE AND COULD NOT PASS THE VERIFY SCREEN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 259497 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |