FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3160022 · Received June 10, 2013

Report

Report Number
2531779-2013-08003
Event Type
Malfunction
Date Received
June 10, 2013
Report Date
June 4, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/28/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT WITH NO EVIDENCE OF PEELING OR OTHER DAMAGE. ALL OF THE KEYPAD BUTTONS WERE FOUND TO BE RESPONDING TO BUTTON PRESSES. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL OF THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS FOUND TO BE FADED AND DISCOLORED. THE DISPLAY SCREEN WAS REMOVED AND REPLACED WITH A TEST SCREEN AND THE DISPLAY RETURNED TO NORMAL. ALSO UNRELATED TO THE COMPLAINT, THE VIBRATION MOTOR PINS WERE FOUND TO NOT MAKE ELECTRICAL CONTACT WITH THE PRINTED CIRCUIT BOARD. ALSO UNRELATED TO THE COMPLAINT, CRACKS IN THE PUMP CASING WERE FOUND AT THE BATTERY COMPARTMENT AND AT THE DISPLAY SCREEN.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260460 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 21 YR