HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE
Report
- Report Number
- 2024168-2013-03632
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- April 15, 2013
- Report Date
- May 17, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- DQX
- PMA / PMN Number
- K103101
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED RESISTANCE WITH THE STENT DELIVERY SYSTEM WAS UNABLE TO BE CONFIRMED. BASED ON A VISUAL, DIMENSIONAL, AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
(B)(4). CONCOMITANT PRODUCTS: STENT DELIVERY SYSTEM: XIENCE XPEDITION; GUIDE CATHETER: 6 FRENCH CORDIS XB 3.5. THE DEVICE WAS RECEIVED. THE INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE XIENCE XPEDITION SDS REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER.
IT WAS REPORTED THAT THE XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) REACHED THE PREDILATED, VERY TORTUOUS, MILDLY CALCIFIED TARGET LESION IN THE DISTAL LEFT ANTERIOR DESCENDING CORONARY ARTERY (LAD). THE PHYSICIAN NOTICED THAT THE BALANCE MIDDLE-WEIGHT (BMW) ELITE GUIDEWIRE MOVED WITH THE SDS, BUT WAS ABLE TO DEPLOY THE STENT WITHOUT ISSUE. THE STENT APPEARED WELL APPOSED TO THE VESSEL. THE SDS WAS DEFLATED, BUT THE SDS COULD NOT BE REMOVED FROM THE GUIDEWIRE; THEREFORE, BOTH DEVICES WERE REMOVED TOGETHER AS A SINGLE UNIT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO REPORTED ADVERSE PATIENT EFFECTS. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260192 | HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE | GUIDE WIRE | DQX | AV-TEMECULA-CT | 2080601 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR |