FDA Adverse Event Injury Summary report: N

ROMEDIC EVA400EE

MDR report key: 3159143 · Received June 3, 2013

Report

Report Number
3005536958-2013-00033
Event Type
Injury
Date Received
June 3, 2013
Date of Event
January 23, 2013
Report Date
June 3, 2013
Manufacturer
HANDICARE AB
Product Code
FSA
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

TWO EMPLOYEES WERE TRANSFERRING A MALE RESIDENT TO A BED USING A PT LIFTING DEVICE. EMPLOYEE 1 WAS GUIDING AND POSITIONING THE RESIDENT WHILE EMPLOYEE 2 WAS PUSHING THE LIFT INTO POSITION, SLIDING THE LIFT LEGS UNDER THE BED. DURING THE TRANSFER, THE LIFT TIPPED FORWARD AND FELL ONTO EMPLOYEE 1, DROPPING THE RESIDENT ONTO THE BED. THE RESIDENT SUSTAINED NO INJURIES BUT EMPLOYEE 1 SUSTAINED A CONTUSION TO THE LEFT SIDE OF HER HEAD, A CONCUSSION, MULTIPLE ABRASIONS AND TWISTED HER BACK. THE LIFT WAS REMOVED FROM THE FLOOR AND TAKEN TO THE FACILITIES MAINTENANCE FOR INSPECTION. THE FACILITIES MAINTENANCE INSPECTED THE UNIT, NO DAMAGE WAS EVIDENT. IT WAS NOTED THAT A SCREW ATTACHING THE LEG SPREAD MECHANISM WAS FOUND TO BE LOOSE. THE FACILITIES MAINTENANCE ORDERED AND REPLACED THE SCREW. THE LIFT WAS RETURNED TO THE FLOOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
243486 ROMEDIC EVA400EE FSA HANDICARE AB EVA400EE

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization