FDA Adverse Event Injury Summary report: N

PAL LIFT

MDR report key: 315883 · Received January 31, 2001

Report

Report Number
MW4002908
Event Type
Injury
Date Received
January 31, 2001
Date of Event
January 22, 2001
Report Date
January 25, 2001
Manufacturer
SUNRISE MACHINE AND TOOL, INC.
Product Code
FSA
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE FACILITY WAS USING A PT LIFT (IN THE LONG TERM CARE FACILITY) IN AN ATTEMPT TO LIFT A PT WHEN THE LIFT CARRIAGE LOCATED AT THE END OF THE LIFT BRACKET AND THREADED INTO THE WEIGHT SENSOR, AND THEN SECURED BY A NUT, CAME OUT OF THE THREADED SENSOR CAUSING THE PT TO FALL TO THE FLOOR. IT APPEARED THAT THE NUT THAT PREVENTS THE CARRIAGE FROM COMING UNTHREADED, DURING THE SWIVEL PROCESS FOR PROPER PLACEMENT OF THE PT, CAME LOOSE FROM MULTIPLE USES, CAUSING THE BOLT TO SPIN FREELY. AFTER A PERIOD OF TIME THE THREADED END OF THE BOLT HAD BECOME SO LOOSE THAT IT CAME OUT ENTIRELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
3826 PAL LIFT PATIENT LIFT FSA SUNRISE MACHINE AND TOOL, INC. PL250SRB *

Patients

Seq Age Sex Outcome Treatment
1 * Life Threatening