FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3157215 · Received June 8, 2013

Report

Report Number
1061932-2013-01098
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
May 11, 2013
Report Date
May 11, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE FOUND THAT THE LOWER SHEATH RESTRICTOR TUBING HAD A HOLE IN IT. THE FSE REPLACED THE TUBING WHICH RESOLVED THE LEAK ISSUE. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IS NOT LIMITED TO PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING A LEAK COMING OUT OF THE RIGHT SIDE OF THE COULTER LH 750 HEMATOLOGY ANALYZER BY THE MIXING CHAMBER WHILE RUNNING CONTROLS. THE VOLUME OF THE LEAK WAS 3 OUNCES AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF THE INCIDENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256721 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 750 NA

Patients

Seq Age Sex Outcome Treatment
1