FDA Adverse Event Injury Summary report: N

VIRTUOSO VR

MDR report key: 3153783 · Received June 8, 2013

Report

Report Number
9614453-2013-01168
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 27, 2013
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS EARLY BATTERY DEPLETION. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED ASA RESULT OF THIS EVENT.

Description of Event or Problem · 1

THE DEVICE WAS LATER EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258476 VIRTUOSO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND D164VWC

Patients

Seq Age Sex Outcome Treatment
1 00048 YR Hospitalization| R 6945-65 IMPLANTABLE TACHY LEAD