FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3153779 · Received June 8, 2013

Report

Report Number
2649622-2013-06217
Event Type
Injury
Date Received
June 8, 2013
Report Date
April 8, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4) IMPLANTABLE PULSE GENERATOR 2007 (B)(6), 5554 IMPLANTABLE PACING LEAD 2007 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE TREND ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS RISING WITH INCREASING LEAD IMPEDANCE TREND FROM APPROXIMATELY 2000 OHMS TO APPROXIMATELY 4000 OHMS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD THRESHOLD HAD GRADUALLY INCREASED TO ALMOST MAXIMUM OUTPUT AT 6 VOLTS AT 0.9 MILLISECONDS. ALSO THE RV LEAD IMPEDANCE WAS GREATER THAN 3000 OHMS. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
255383 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00075 YR Hospitalization| R