FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 3153740
·
Received June 8, 2013
Report
- Report Number
- 2649622-2013-06188
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- March 14, 2013
- Report Date
- March 15, 2013
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. HOWEVER, THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AT THE IMPLANT PROCEDURE, THE LEAD HELIX WOULD NOT FULLY RETRACT DURING LEAD REPOSITIONING. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 255074 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00088 YR |