FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 3153736
·
Received June 8, 2013
Report
- Report Number
- 2649622-2013-06193
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- January 17, 2013
- Report Date
- March 29, 2013
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY:THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE HELIX OF THE LEAD WAS EXTRINSICALLY COMPRESSED AND THE GUIDE TOOTH OF THE LEAD WAS EXTRINSICALLY DAMAGED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE HELIX OF THE RIGHT VENTRICULAR LEAD WOULD NOT DEPLOY. A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 257677 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00078 YR |