FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3153675 · Received June 8, 2013

Report

Report Number
2183613-2013-00619
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
March 22, 2013
Report Date
April 5, 2013
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE CUSTOMER COMMENT THAT THE BATTERY DOOR WAS STICKING, IT WAS ATTRIBUTED TO THE UPPER CASE, BATTERY RELEASE AND BATTERY DRAWER BEING CONTAMINATED AND THE BATTERY DRAWER ALSO BEING BROKEN. ANALYSIS ALSO FOUND THE LOWER CASE BROKEN AND THE BATTERY CONTACTS COMPRESSED. (B)(4).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BATTERY DOOR OF THE EXTERNAL PULSE GENERATOR WOULD STICK AND WAS NOT ABLE TO BE OPENED. THE GENERATOR WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BATTERY DOOR OF THE EXTERNAL PULSE GENERATOR WOULD STICK AND WAS NOT ABLE TO BE OPENED. THE GENERATOR WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256457 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1