PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2183613-2013-00619
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- March 22, 2013
- Report Date
- April 5, 2013
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE CUSTOMER COMMENT THAT THE BATTERY DOOR WAS STICKING, IT WAS ATTRIBUTED TO THE UPPER CASE, BATTERY RELEASE AND BATTERY DRAWER BEING CONTAMINATED AND THE BATTERY DRAWER ALSO BEING BROKEN. ANALYSIS ALSO FOUND THE LOWER CASE BROKEN AND THE BATTERY CONTACTS COMPRESSED. (B)(4).
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(6). (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE BATTERY DOOR OF THE EXTERNAL PULSE GENERATOR WOULD STICK AND WAS NOT ABLE TO BE OPENED. THE GENERATOR WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE BATTERY DOOR OF THE EXTERNAL PULSE GENERATOR WOULD STICK AND WAS NOT ABLE TO BE OPENED. THE GENERATOR WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256457 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MILACA, INC. | 5388 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |