FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY PLUS

MDR report key: 3153674 · Received June 8, 2013

Report

Report Number
2649622-2013-06153
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
March 26, 2013
Report Date
March 26, 2013
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. HOWEVER, THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. THE TIP SEAL ON THE DISTAL ELECTRODE OF THE LEAD SHOWED EVIDENCE OF BLOOD INGRESSION. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE LEAD WOULD NOT STAY IN THE LATERAL BRANCH AND KEPT DISLODGING. THE LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
255923 ATTAIN ABILITY PLUS DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 429688

Patients

Seq Age Sex Outcome Treatment
1 00092 YR