FDA Adverse Event Injury Summary report: N

ADAPTA DR

MDR report key: 3153673 · Received June 8, 2013

Report

Report Number
3004209178-2013-08935
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 26, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY- THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 4092-52 IMPLANTABLE PACING LEAD (B)(6) 2010; 4574-45 IMPLANTABLE PACING LEAD (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE MAY BE PREMATURE BATTERY DEPLETION. IT WAS ALSO REPORTED PER FLUOROSCOPY ON THE DAY OF THE GENERATOR ANDLEAD CHANGE THE PHYSICIAN NOTED THAT THE TIP OF THE ATRIAL LEAD HAD DROPPED DOWN TO JUST ABOVE THE ANNULUS. IN ADDITION THERE WAS INTERMITTENT CAPTURE UNIPOLAR AND BIPOLAR AT 8V/1 MSEC. THE DEVICE WAS EXPLANTED AND REPLACED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254100 ADAPTA DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADDR01

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R