FDA Adverse Event Injury Summary report: N

MAXIMO II DR

MDR report key: 3153633 · Received June 8, 2013

Report

Report Number
3004209178-2013-08930
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 25, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 82% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. ADDITIONALLY, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED AND ANALYSIS REVEALED THE DEVICE WAS PRE/APPROACHING RECOMMENDED REPLACEMENT (RRT). CONCOMITANT PRODUCT: 6949, IMPLANTABLE TACHY LEAD, (B)(6) 2005. (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254504 MAXIMO II DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D284DRG

Patients

Seq Age Sex Outcome Treatment
1 00088 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD