PROTECTA DR
Report
- Report Number
- 3004209178-2013-08927
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- April 12, 2013
- Report Date
- April 12, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 5076 IMPLANTABLE PACING LEAD - (B)(6) 2007, 5076 IMPLANTABLE PACING LEAD - (B)(6) 2009. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE RETURNED DEVICE INDICATED THE SET SCREW WAS FOUND IN THE CONNECTOR BORE.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE DEVICE EXHIBITED OVERSENSING AND AN OUTPUT FAILURE. THE DEVICE WAS DETERMINED TO EXHIBIT A SETSCREW ISSUE. THE DEVICE WAS NOT USED AND ANOTHER DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE DEVICE EXHIBITED OVERSENSING AND AN OUTPUT FAILURE. THE DEVICE WAS DETERMINED TO EXHIBIT A SETSCREW ISSUE. THE DEVICE WAS NOT USED AND ANOTHER DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 258841 | PROTECTA DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D334DRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00075 YR | 6949 IMPLANTABLE TACHY LEAD |