SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2013-06090
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- April 3, 2013
- Report Date
- April 3, 2013
- Manufacturer
- MPRI
- Product Code
- NVY
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS FOUND THAT THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT AT THE IMPLANT PROCEDURE, THE LEAD HELIX WOULD NOT DEPLOY. THE LEAD WAS REMOVED AND TESTED OUTSIDE OF THE PATIENT. AFTER 30 TURNS THE HELIX ¿JUMPED¿ OUT OF THE HOUSING. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 257399 | SPRINT QUATTRO SECURE | DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES | NVY | MPRI | 6947M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR |