FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3153537 · Received June 8, 2013

Report

Report Number
2649622-2013-06074
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 21, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT ID: E2DR01AA IMPLANTABLE PULSE GENERATOR (IPG), (B)(6) 2004; 5076 IMPLANTABLE PACING LEAD, (B)(6) 2004. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD BIPOLAR IMPEDANCE HAS BEEN GRADUALLY INCREASING AND WAS NOW HIGH. THE CAPTURE THRESHOLD HAS ALSO INCREASED. CRUSH FRACTURE WAS SUSPECTED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258144 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R