FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3153528 · Received June 8, 2013

Report

Report Number
2649622-2013-06063
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 19, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4). CONCOMITANT PRODUCT: 4076 IMPLANTABLE PACING LEAD (B)(6) 2006.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD SHOWED 9 LOW LEAD IMPEDANCE MEASUREMENTS AND TRIGGERED A POLARITY SWITCH. PARAMETERS ON THE LEAD WERE REPROGRAMMED AND IT REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258137 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention ADDR01 IMPLANTABLE PULSE GENERATOR (IPG)