FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3153509 · Received June 8, 2013

Report

Report Number
2649622-2013-06070
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
April 1, 2013
Report Date
April 1, 2013
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE EXPOSED SVC (SUPERIOR VENA CAVA) DEFIBRILLATION COIL BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. IT WAS NOTED THAT VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, WHILE INSERTING THE RIGHT VENTRICULAR (RV) LEAD WITH MUCH TORTUOSITY UNDER FLUOROSCOPY A SEPARATION IN THE SUPERIOR VENA CAVA COIL OF THE RV LEAD WAS NOTED. THE RV LEAD WAS REMOVED AND A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257743 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00079 YR