ATTAIN ABILITY STRAIGHT
Report
- Report Number
- 2649622-2013-06064
- Event Type
- Injury
- Date Received
- June 8, 2013
- Date of Event
- March 25, 2013
- Report Date
- March 26, 2013
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. IT WAS NOTED THAT THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THE INNER TUBING OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE ANALYST COMMENTED THAT DUE TO THE SUBSTANTIAL AMOUNT OF BLOOD INGRESS AND THE SHORTED CONDITION OF THE CONDUCTORS, IT IS LIKELY THAT THE DAMAGE TO THE INSULATION OCCURRED DURING THE INITIAL IMPLANT. (B)(4).
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 1346T COMPETITOR IMPLANTABLE PACING LEAD: (B)(6) 1998. 1388T COMPETITOR IMPLANTABLE PACING LEAD: (B)(6) 1998. (B)(4).
IT WAS REPORTED THAT POST IMPLANT OF THE LEFT VENTRICULAR (LV) LEAD THERE WAS NOTED HIGH THRESHOLDS AS WELL AS DIAPHRAGMATIC STIMULATION. THEREFORE THE LV LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256695 | ATTAIN ABILITY STRAIGHT | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 439688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00089 YR | Hospitalization| R | C4TR01 IMPLANTABLE PACEMAKER CARDIO/DEFIB |