SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2013-06060
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- March 28, 2013
- Report Date
- March 28, 2013
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE RIGHT VENTRICULAR LEAD R-WAVES DECREASED AFTER SEVERAL REPOSITIONINGS. THE PHYSICIAN REMOVED AND CLEANED THE LEAD, BUT THE LOW R-WAVES CONTINUED. A NEW LEAD WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256505 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00069 YR |